WAC 246-290-694
Monitoring for unfiltered systems. (1)
Source coliform monitoring for systems without a limited
alternative to filtration.
(a) The purveyor shall ensure that source water samples
of each surface or GWI source are representative and:
(i) Collected before the first point of disinfectant
application; and
(ii) Analyzed for fecal coliform density in accordance
with methods acceptable to the department.
(b) The purveyor shall ensure source samples are
collected for fecal coliform analysis each week the system
serves water to the public based on the following schedule:
|
Population
Served |
Minimum
Number/week* |
| 25 |
- |
500 |
1 |
| 501 |
- |
3,300 |
2 |
| 3,301 |
- |
10,000 |
3 |
| 10,001 |
- |
25,000 |
4 |
| >25,000 |
|
5 |
*Must be taken on separate days.
(c) Each day the system serves water to the public and
the turbidity of the source water exceeds 1.0 NTU, the
purveyor shall ensure one representative source water sample
is collected before the first point of disinfectant
application and analyzed for fecal coliform density. This
sample shall count toward the weekly source coliform sampling
requirement.
(d) The purveyor using a surface water or GWI source and
that meets the criteria to remain unfiltered under WAC 246-290-690, shall collect at least one sample near the first
service connection each day the turbidity level of the source
water, measured as specified under WAC 246-290-694, exceeds 1
NTU. This sample must be analyzed for the presence of total
coliform. When one or more turbidity measurements in any day
exceed 1 NTU, the system must collect this coliform sample
within twenty-four hours of the first exceedance, unless the
department determines that the system, for logistical reasons
outside the system's control, cannot have the sample analyzed
within thirty hours of collection. Sample results from this
coliform monitoring must be included in determining compliance
with the MCL for total coliforms under WAC 246-290-310(2).
(e) A purveyor shall not be considered in violation of
(c) of this subsection, if the purveyor demonstrates to the
department's satisfaction that, for valid logistical reasons
outside the purveyor's control, the additional fecal coliform
sample could not be analyzed within a time frame acceptable to
the department.
(2) Source coliform monitoring for systems with a limited
alternative to filtration.
(a) The purveyor shall ensure that source water samples
of each surface or GWI source are:
(i) Collected before the first point of primary
disinfection; and
(ii) Analyzed for fecal coliform density in accordance
with methods acceptable to the department.
(b) At a minimum, the purveyor shall ensure source
samples are collected for fecal coliform analysis at a
frequency equal to ten percent the number of routine coliform
samples collected within the distribution system each month
under WAC 246-290-300, or once per calendar month, whichever
is greater, up to a maximum of one sample per day.
(3) Coliform monitoring at entry to distribution for
systems without a limited alternative to filtration.
(a) The purveyor shall collect and have analyzed one
coliform sample at the entry point to the distribution system
each day that a routine or repeat coliform sample is collected
within the distribution system under WAC 246-290-300(3) or
246-290-320(2), respectively.
(b) The purveyor shall use the results of the coliform
monitoring at entry to distribution along with inactivation
ratio monitoring results to demonstrate the adequacy of source
treatment.
(4) Source turbidity monitoring for systems without a
limited alternative to filtration.
(a) The purveyor shall continuously monitor and record
turbidity:
(i) On representative source water samples before the
first point of primary disinfectant application; and
(ii) In accordance with the analytical techniques in WAC 246-290-638.
(b) If source water turbidity is not the same as the
turbidity of water delivered to consumers, the purveyor shall
continuously monitor and record turbidity of water delivered.
(5) Source turbidity monitoring for systems with a
limited alternative to filtration. The purveyor shall:
(a) Continuously monitor turbidity on representative
source samples before the first point of primary disinfection
application;
(b) Record continuous turbidity measurements at equal
intervals, of at least four hours, in accordance with a
department-approved sampling schedule; and
(c) Conduct monitoring in accordance with the analytical
techniques under WAC 246-290-638.
(6) Monitoring the level of inactivation.
(a) Each day the system is in operation, the purveyor
shall determine the total level of inactivation of Giardia
lamblia cysts, viruses, and, if providing a limited
alternative to filtration, any other pathogenic organisms of
health concern including Cryptosporidium oocysts, achieved
through disinfection.
(b) At least once per day, the purveyor shall monitor the
following parameters to determine the total inactivation ratio
achieved through disinfection:
(i) Temperature of the disinfected water at each residual
disinfectant concentration sampling point used for CT
calculations; and
(ii) If using chlorine, pH of the disinfected water at
each chlorine residual disinfectant concentration sampling
point used for CT calculations.
(c) Each day during peak hourly flow, the purveyor shall:
(i) Determine disinfectant contact time, T, to the point
at which C is measured; and
(ii) Measure the residual disinfectant concentration, C,
of the water at the point for which T is calculated. The C
measurement point must be before or at the first consumer.
(7) Monitoring the residual disinfectant concentration
entering the distribution system for either unfiltered
systems, or systems using a limited alternative to filtration.
(a) Systems serving more than thirty-three hundred
people.
(i) The purveyor shall continuously monitor and record
the residual disinfectant concentration of water entering the
distribution system and report the lowest value each day.
(ii) If the continuous monitoring equipment fails, the
purveyor shall measure the residual disinfectant concentration
on grab samples collected at least every four hours at the
entry to the distribution system while the equipment is being
repaired or replaced. The purveyor shall have continuous
monitoring equipment back on-line within five working days
following failure.
(b) Systems serving thirty-three hundred or less people.
(i) The purveyor shall collect grab samples or use
continuous monitoring and recording to measure the residual
disinfectant concentration entering the distribution system.
(ii) A purveyor choosing to take grab samples shall
collect:
(A) Samples at the following minimum frequencies:
| |
Population
Served |
Number/day |
| 25 |
- |
500 |
1 |
| 501 |
- |
1,000 |
2 |
| 1,001 |
- |
2,500 |
3 |
| 2,501 |
- |
3,300 |
4 |
(B) At least one of the grab samples at peak hourly flow
based on historical flows for the system; and
(C) The remaining sample or samples at intervals evenly
spaced over the time the system is disinfecting water that
will be delivered to the public.
(iii) When grab samples are collected and the residual
disinfectant concentration at the entry to distribution falls
below 0.2 mg/L, the purveyor shall collect a grab sample every
four hours until the residual disinfectant concentration is
0.2 mg/L or more.
(8) Monitoring residual disinfectant concentration within
the distribution system for either unfiltered systems, or
systems using a limited alternative to filtration.
(a) The purveyor shall measure the residual disinfectant
concentration within the distribution system at the same time
and location that a routine or repeat coliform sample is
collected under WAC 246-290-300(3) or 246-290-320(2) or once
per day, whichever is greater.
(b) The purveyor of a system that purchases completely
treated surface or GWI water as determined by the department
shall comply with the requirements of (a) of this subsection
or as otherwise directed by the department under WAC 246-290-300(2). At a minimum, the purveyor shall measure the
residual disinfectant concentration within the distribution
system at the same time and location that a routine or repeat
coliform sample is collected under WAC 246-290-300(3) or
246-290-320(2).
(c) The purveyor may measure HPC within the distribution
system in lieu of measuring the residual disinfectant
concentration under this subsection.
[Statutory Authority: RCW 43.20.050. 09-21-045, §
246-290-694, filed 10/13/09, effective 1/4/10. Statutory
Authority: RCW 70.119A.180 and 43.20.050. 08-03-061, §
246-290-694, filed 1/14/08, effective 2/14/08. Statutory
Authority: RCW 43.20.050 (2) and (3) and 70.119A.080. 03-08-037, § 246-290-694, filed 3/27/03, effective 4/27/03. Statutory Authority: RCW 43.02.050 [43.20.050]. 99-07-021, §
246-290-694, filed 3/9/99, effective 4/9/99. Statutory
Authority: RCW 43.20.050. 94-14-001, § 246-290-694, filed
6/22/94, effective 7/23/94; 93-08-011 (Order 352B), §
246-290-694, filed 3/25/93, effective 4/25/93.]