WAC 246-290-460
Fluoridation of drinking water. (1)
Purveyors shall obtain written department approval of
fluoridation treatment facilities before placing them in
service.
(2) Where fluoridation is practiced, purveyors shall
maintain fluoride concentrations in the range 0.8 through 1.3
mg/L throughout the distribution system.
(3) Where fluoridation is practiced, purveyors shall take
the following actions to ensure that concentrations remain at
optimal levels and that fluoridation facilities and monitoring
equipment are operating properly:
(a) Daily monitoring.
(i) Take daily monitoring samples for each point of
fluoride addition and analyze the fluoride concentration.
Samples must be taken downstream from each fluoride injection
point at the first sample tap where adequate mixing has
occurred.
(ii) Record the results of daily analyses in a monthly
report format acceptable to the department. A report must be
made for each point of fluoride addition.
(iii) Submit monthly monitoring reports to the department
within the first ten days of the month following the month in
which the samples were collected.
(b) Monthly split sampling.
(i) Take a monthly split sample at the same location
where routine daily monitoring samples are taken. A monthly
split sample must be taken for each point of fluoride
addition.
(ii) Analyze a portion of the sample and record the
results on the lab sample submittal form and on the monthly
report form.
(iii) Forward the remainder of the sample, along with the
completed sample form to the state public health laboratory,
or other state-certified laboratory, for fluoride analysis.
(iv) If a split sample is found by the certified lab to
be:
(A) Not within the range of 0.8 to 1.3 mg/l, the
purveyor's fluoridation process shall be considered out of
compliance.
(B) Differing by more than 0.30 mg/l from the purveyor's
analytical result, the purveyor's fluoride testing shall be
considered out of control.
(4) Purveyors shall conduct analyses prescribed in
subsection (3) of this section in accordance with procedures
listed in the most recent edition of Standard Methods for the
Examination of Water and Wastewater.
(5) The purveyor may be required by the department to
increase the frequency, and/or change the location of sampling
prescribed in subsection (3) of this section to ensure the
adequacy and consistency of fluoridation.
[Statutory Authority: RCW 43.02.050 [43.20.050]. 99-07-021,
§ 246-290-460, filed 3/9/99, effective 4/9/99. Statutory
Authority: RCW 43.20.050. 91-02-051 (Order 124B), recodified
as § 246-290-460, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 34.04.045. 88-05-057 (Order 307), §
248-54-235, filed 2/17/88. Statutory Authority: RCW 43.20.050. 83-19-002 (Order 266), § 248-54-235, filed
9/8/83.]