WAC 246-240-623
Records of periodic spot-checks for
gamma stereotactic radiosurgery units. (1) A licensee shall
retain a record of each spot-check for gamma stereotactic
radiosurgery units required by WAC 246-240-384 for three
years.
(2) The record must include:
(a) The date of the spot-check;
(b) The manufacturer's name, model number, and serial
number for the gamma stereotactic radiosurgery unit and the
instrument used to measure the output of the unit;
(c) An assessment of timer linearity and accuracy;
(d) The calculated on-off error;
(e) A determination of trunnion centricity;
(f) The difference between the anticipated output and the
measured output;
(g) An assessment of source output against computer
calculations;
(h) Notations indicating the operability of radiation
monitors, helmet microswitches, emergency timing circuits,
emergency off buttons, electrical interlocks, source exposure
indicator lights, viewing and intercom systems, timer
termination, treatment table retraction mechanism, and
stereotactic frames and localizing devices (trunnions); and
(i) The name of the individual who performed the periodic
spot-check and the signature of the authorized medical
physicist who reviewed the record of the spot-check.
(3) A licensee shall retain a copy of the procedures
required by WAC 246-240-384(2) until the licensee no longer
possesses the gamma stereotactic radiosurgery unit.
[Statutory Authority: RCW 70.98.050. 06-05-019, §