WAC 246-240-396
Therapy-related computer systems. The
licensee shall perform acceptance testing on the treatment
planning system of therapy-related computer systems in
accordance with published protocols accepted by nationally
recognized bodies. At a minimum, the acceptance testing must
include, as applicable, verification of:
(1) The source-specific input parameters required by the
dose calculation algorithm;
(2) The accuracy of dose, dwell time, and treatment time
calculations at representative points;
(3) The accuracy of isodose plots and graphic displays;
(4) The accuracy of the software used to determine sealed
source positions from radiographic images; and
(5) The accuracy of electronic transfer of the treatment
delivery parameters to the treatment delivery unit from the
treatment planning system.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-396, filed 2/6/06, effective 3/9/06.]