WAC 246-240-387
Additional technical requirements for
mobile remote afterloader units. (1) A licensee providing
mobile remote afterloader service shall:
(a) Check survey instruments before medical use at each
address of use or on each day of use, whichever is more
frequent; and
(b) Account for all sources before departure from a
client's address of use.
(2) In addition to the periodic spot-checks required by
WAC 246-240-381, a licensee authorized to use mobile
afterloaders for medical use shall perform checks on each
remote afterloader unit before use at each address of use. At
a minimum, checks must be made to verify the operation of:
(a) Electrical interlocks on treatment area access
points;
(b) Source exposure indicator lights on the remote
afterloader unit, on the control console, and in the facility;
(c) Viewing and intercom systems;
(d) Applicators, source transfer tubes, and transfer
tube-applicator interfaces;
(e) Radiation monitors used to indicate room exposures;
(f) Source positioning (accuracy); and
(g) Radiation monitors used to indicate whether the
source has returned to a safe shielded position.
(3) In addition to the requirements for checks in
subsection (2) of this section, a licensee shall ensure
overall proper operation of the remote afterloader unit by
conducting a simulated cycle of treatment before use at each
address of use.
(4) If the results of the checks required in subsection
(2) of this section indicate the malfunction of any system, a
licensee shall lock the control console in the off position
and not use the unit except as may be necessary to repair,
replace, or check the malfunctioning system.
(5) A licensee shall retain a record of each check
required by subsection (2) of this section in accordance with
WAC 246-240-626.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-387, filed 2/6/06, effective 3/9/06.]