WAC 246-240-384
Periodic spot-checks for gamma
stereotactic radiosurgery units. (1) A licensee authorized to
use a gamma stereotactic radiosurgery unit for medical use
shall perform spot-checks of each gamma stereotactic
radiosurgery facility and on each unit:
(a) Monthly;
(b) Before the first use of the unit on a given day; and
(c) After each source installation.
(2) A licensee shall:
(a) Perform the measurements required by subsection (1)
of this section in accordance with written procedures
established by the authorized medical physicist. That
individual need not actually perform the spot check
measurements.
(b) Have the authorized medical physicist review the
results of each spot-check within fifteen days. The
authorized medical physicist shall notify the licensee as soon
as possible in writing of the results of each spot-check.
(3) To satisfy the requirements of subsection (1)(a) of
this section, spot-checks must, at a minimum:
(a) Assure proper operation of:
(i) Treatment table retraction mechanism, using backup
battery power or hydraulic backups with the unit off;
(ii) Helmet microswitches;
(iii) Emergency timing circuits; and
(iv) Stereotactic frames and localizing devices
(trunnions).
(b) Determine:
(i) The output for one typical set of operating
conditions measured with the dosimetry system described in WAC 246-240-366(2);
(ii) The difference between the measurement made in
(b)(i) of this subsection and the anticipated output,
expressed as a percentage of the anticipated output (i.e., the
value obtained at last full calibration corrected
mathematically for physical decay);
(iii) Source output against computer calculation;
(iv) Timer accuracy and linearity over the range of use;
(v) On-off error; and
(vi) Trunnion centricity.
(4) To satisfy the requirements of subsection (1)(b) and
(c) of this section, spot-checks must assure proper operation
of:
(a) Electrical interlocks at each gamma stereotactic
radiosurgery room entrance;
(b) Source exposure indicator lights on the gamma
stereotactic radiosurgery unit, on the control console, and in
the facility;
(c) Viewing and intercom systems;
(d) Timer termination;
(e) Radiation monitors used to indicate room exposures;
and
(f) Emergency off buttons.
(5) A licensee shall arrange for the repair of any system
identified in subsection (3) of this section that is not
operating properly as soon as possible.
(6) If the results of the checks required in subsection
(4) of this section indicate the malfunction of any system, a
licensee shall lock the control console in the off position
and not use the unit except as may be necessary to repair,
replace, or check the malfunctioning system.
(7) A licensee shall retain a record of each check
required by subsections (3) and (4) of this section and a copy
of the procedures required by subsection (2) of this section
in accordance with WAC 246-240-623.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-384, filed 2/6/06, effective 3/9/06.]