WAC 246-240-378
Periodic spot-checks for teletherapy
units. (1) A licensee authorized to use teletherapy units for
medical use shall perform output spot-checks on each
teletherapy unit once in each calendar month that include
determination of:
(a) Timer accuracy, and timer linearity over the range of
use;
(b) On-off error;
(c) The coincidence of the radiation field and the field
indicated by the light beam localizing device;
(d) The accuracy of all distance measuring and
localization devices used for medical use;
(e) The output for one typical set of operating
conditions measured with the dosimetry system described in WAC 246-240-366(2); and
(f) The difference between the measurement made in (e) of
this subsection and the anticipated output, expressed as a
percentage of the anticipated output (i.e., the value obtained
at last full calibration corrected mathematically for physical
decay).
(2) A licensee shall perform measurements required by
subsection (1) of this section in accordance with written
procedures established by the authorized medical physicist.
That individual need not actually perform the spot-check
measurements.
(3) A licensee shall have the authorized medical
physicist review the results of each spot-check within fifteen
days. The authorized medical physicist shall notify the
licensee as soon as possible in writing of the results of each
spot-check.
(4) A licensee authorized to use a teletherapy unit for
medical use shall perform safety spot-checks of each
teletherapy facility once in each calendar month and after
each source installation to assure proper operation of:
(a) Electrical interlocks at each teletherapy room
entrance;
(b) Electrical or mechanical stops installed for the
purpose of limiting use of the primary beam of radiation
(restriction of source housing angulation or elevation,
carriage or stand travel and operation of the beam on-off
mechanism);
(c) Source exposure indicator lights on the teletherapy
unit, on the control console, and in the facility;
(d) Viewing and intercom systems;
(e) Treatment room doors from inside and outside the
treatment room; and
(f) Electrically assisted treatment room doors with the
teletherapy unit electrical power turned off.
(5) If the results of the checks required in subsection
(4) of this section indicate the malfunction of any system, a
licensee shall lock the control console in the off position
and not use the unit except as may be necessary to repair,
replace, or check the malfunctioning system.
(6) A licensee shall retain a record of each spot-check
required by subsections (1) and (4) of this section, and a
copy of the procedures required by subsection (2) of this
section, in accordance with WAC 246-240-617.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-378, filed 2/6/06, effective 3/9/06.]