WAC 246-240-372
Full calibration measurements on remote
afterloader units. (1) A licensee authorized to use a remote
afterloader unit for medical use shall perform full
calibration measurements on each unit:
(a) Before the first medical use of the unit;
(b) Before medical use under the following conditions:
(i) Following replacement of the source or following
reinstallation of the unit in a new location outside the
facility; and
(ii) Following any repair of the unit that includes
removal of the source or major repair of the components
associated with the source exposure assembly; and
(c) At intervals not exceeding one calendar quarter for
high dose-rate, medium dose-rate, and pulsed dose-rate remote
afterloader units with sources whose half-life exceeds
seventy-five days; and
(d) At intervals not exceeding one year for low dose-rate
remote afterloader units.
(2) To satisfy the requirement of subsection (1) of this
section, full calibration measurements must include, as
applicable, determination of:
(a) The output within ±5 percent;
(b) Source positioning accuracy to within ±1 millimeter;
(c) Source retraction with backup battery upon power
failure;
(d) Length of the source transfer tubes;
(e) Timer accuracy and linearity over the typical range
of use;
(f) Length of the applicators; and
(g) Function of the source transfer tubes, applicators,
and transfer tube-applicator interfaces.
(3) A licensee shall use the dosimetry system described
in WAC 246-240-366(1) to measure the output.
(4) A licensee shall make full calibration measurements
required by subsection (1) of this section in accordance with
published protocols accepted by nationally recognized bodies.
(5) In addition to the requirements for full calibrations
for low dose-rate remote afterloader units in subsection (2)
of this section, a licensee shall perform an autoradiograph of
the source(s) to verify inventory and source(s) arrangement at
intervals not exceeding one calendar quarter.
(6) For low dose-rate remote afterloader units, a
licensee may use measurements provided by the source
manufacturer that are made in accordance with subsections (1)
through (5) of this section.
(7) A licensee shall mathematically correct the outputs
determined in subsection (2)(a) of this section for physical
decay at intervals consistent with one percent physical decay.
(8) Full calibration measurements required by subsection
(1) of this section and physical decay corrections required by
subsection (7) of this section must be performed by the
authorized medical physicist.
(9) A licensee shall retain a record of each calibration
in accordance with WAC 246-240-614.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-372, filed 2/6/06, effective 3/9/06.]