WAC 246-240-366
Dosimetry equipment. (1) Except for low
dose-rate remote afterloader sources where the source output
or activity is determined by the manufacturer, a licensee
shall have a calibrated dosimetry system available for use.
To satisfy this requirement, one of the following two
conditions must be met:
(a) The system must have been calibrated using a system
or source traceable to the National Institute of Science and
Technology (NIST) and published protocols accepted by
nationally recognized bodies; or by a calibration laboratory
accredited by the American Association of Physicists in
Medicine (AAPM). The calibration must have been performed
within the previous two years and after any servicing that may
have affected system calibration; or
(b) The system must have been calibrated within the
previous four years. Eighteen to thirty months after that
calibration, the system must have been intercompared with
another dosimetry system that was calibrated within the past
twenty-four months by NIST or by a calibration laboratory
accredited by the AAPM. The results of the intercomparison
must indicate that the calibration factor of the licensee's
system had not changed by more than two percent. The licensee
may not use the intercomparison result to change the
calibration factor. When intercomparing dosimetry systems to
be used for calibrating sealed sources for therapeutic units,
the licensee shall use a comparable unit with beam attenuators
or collimators, as applicable, and sources of the same
radionuclide as the source used at the licensee's facility.
(2) The licensee shall have a dosimetry system available
for use for spot-check output measurements, if applicable. To
satisfy this requirement, the system may be compared with a
system that has been calibrated in accordance with subsection
(1) of this section. This comparison must have been performed
within the previous year and after each servicing that may
have affected system calibration. The spot-check system may
be the same system used to meet the requirement in subsection
(1) of this section.
(3) The licensee shall retain a record of each
calibration, intercomparison, and comparison in accordance
with WAC 246-240-611.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-366, filed 2/6/06, effective 3/9/06.]