WAC 246-240-357
Installation, maintenance, adjustment,
and repair. (1) Only a person specifically licensed by the
department, the U.S. NRC, or an agreement state shall install,
maintain, adjust, or repair a remote afterloader unit,
teletherapy unit, or gamma stereotactic radiosurgery unit that
involves work on the source(s) shielding, the source(s)
driving unit, or other electronic or mechanical component that
could expose the source(s), reduce the shielding around the
source(s), or compromise the radiation safety of the unit or
the source(s).
(2) Except for low dose-rate remote afterloader units,
only a person specifically licensed by the department, the
U.S. NRC, or an agreement state shall install, replace,
relocate, or remove a sealed source or source contained in
other remote afterloader units, teletherapy units, or gamma
stereotactic radiosurgery units.
(3) For a low dose-rate remote afterloader unit, only a
person specifically licensed by the department, the U.S. NRC,
or an agreement state or an authorized medical physicist shall
install, replace, relocate, or remove a sealed source(s)
contained in the unit.
(4) A licensee shall retain a record of the installation,
maintenance, adjustment, and repair of remote afterloader
units, teletherapy units, and gamma stereotactic radiosurgery
units in accordance with WAC 246-240-605.
[Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-357, filed 2/6/06, effective 3/9/06.]