WAC 246-240-275   Therapy-related computer systems.  The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

     (1) The source-specific input parameters required by the dose calculation algorithm;

     (2) The accuracy of dose, dwell time, and treatment time calculations at representative points;

     (3) The accuracy of isodose plots and graphic displays; and

     (4) The accuracy of the software used to determine sealed source positions from radiographic images.



[Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-275, filed 2/6/06, effective 3/9/06.]