WAC 246-240-269
Calibration measurements of
brachytherapy sources. (1) Before the first medical use of a
brachytherapy source on or after March 9, 2006,
a licensee shall have:
(a) Determined the source output or activity using a
dosimetry system that meets the requirements of WAC 246-240-366(1);
(b) Determined source positioning accuracy within
applicators; and
(c) Used published protocols currently accepted by
nationally recognized bodies to meet the requirements of (a)
and (b) of this subsection.
(2) A licensee may use measurements provided by the
source manufacturer or by a calibration laboratory accredited
by the American Association of Physicists in Medicine that are
made in accordance with subsection (1) of this section.
(3) A licensee shall mathematically correct the outputs
or activities determined in subsection (1) of this section for
physical decay at intervals consistent with one percent
physical decay.
(4) A licensee shall retain a record of each calibration
in accordance with WAC 246-240-599.
[Statutory Authority: RCW 70.98.050. 07-14-131, §
246-240-269, filed 7/3/07, effective 8/3/07; 06-05-019, §