WAC 246-240-160
Permissible molybdenum-99 concentration. (1) A licensee may not administer to humans a
radiopharmaceutical that contains more than:
(a) 5.55 kilobecquerel of molybdenum-99 per 37
megabecquerel of technetium-99m (0.15 microcurie of
molybdenum-99 per millicurie of technetium-99m); or
(b) 0.02 kilobecquerel of strontium-82 per megabecquerel
of rubidium-82 chloride injection, (0.02 microcurie of
strontium-82 per millicurie of rubidium-82 chloride); or
(c) 0.2 kilobecquerel of strontium-85 per megabecquerel
of rubidium-82 chloride injection (0.2 microcurie of
strontium-85 per millicurie of rubidium-82).
(2) A licensee that uses molybdenum-99/technetium-99m
generators for preparing a technetium-99m radiopharmaceutical
shall measure the molybdenum-99 concentration of the first
eluate after receipt of a generator to demonstrate compliance
with subsection (1) of this section.
(3) A licensee that uses a strontium-82/rubidium-82
generator for preparing a rubidium-82 radiopharmaceutical
shall, before the first patient use of the day, measure the
concentration of strontium-82 and strontium-85 to demonstrate
compliance with subsection (1)(a) of this section.
(4) If a licensee is required to measure the
molybdenum-99 concentration, or strontium-82 and strontium-85
concentrations the licensee shall retain a record of each
measurement in accordance with WAC 246-240-587.
[Statutory Authority: RCW 70.98.050 and 70.98.080. 09-06-003, § 246-240-160, filed 2/18/09, effective 3/21/09. Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-160, filed 2/6/06, effective 3/9/06.]