WAC 246-240-113
Requirements for possession of sealed
sources and brachytherapy sources. (1) A licensee in
possession of any sealed source or brachytherapy source shall
follow the radiation safety and handling instructions supplied
by the manufacturer.
(2) A licensee in possession of a sealed source shall:
(a) Test the source for leakage before its first use
unless the licensee has a certificate from the supplier
indicating that the source was tested within six months before
transfer to the licensee; and
(b) Test the source for leakage at intervals not to
exceed six months or at other intervals approved by the
department, the U.S. NRC, or an agreement state in the sealed
source and device registry.
(3) To satisfy the leak test requirements of this
section, the licensee shall ensure the sample is analyzed by
such method that the leak test can detect the presence of 185
becquerels (0.005 microcuries) of radioactive material in the
sample.
(4) A licensee shall retain leak test records in
accordance with WAC 246-240-572(1).
(5) If the leak test reveals the presence of 185
becquerels (0.005 microcuries) or more of removable
contamination, the licensee shall:
(a) Immediately withdraw the sealed source from use and
store, dispose, or cause it to be repaired in accordance with
the requirements in chapters 246-221 and 246-232 WAC; and
(b) File a report within five days of the leak test in
accordance with WAC 246-240-657.
(6) A licensee need not perform a leak test on the
following sources:
(a) Sources containing only radioactive material with a
half-life of less than thirty days;
(b) Sources containing only radioactive material as a
gas;
(c) Sources containing 3.7 megabecquerels (100
microcuries) or less of beta- or gamma-emitting material or
0.37 megabecquerels (10 microcuries) or less of alpha-emitting
material;
(d) Seeds of iridium-192 encased in nylon ribbon; and
(e) Sources stored and not being used. However, the
licensee shall test each source for leakage before any use or
transfer unless it has been leak tested within six months
before the date of use or transfer.
(7) A licensee in possession of sealed sources or
brachytherapy sources, except for gamma stereotactic
radiosurgery sources, shall conduct a physical inventory of
all the sources in its possession at intervals not to exceed
six months. The licensee shall retain each inventory record
in accordance with WAC 246-240-572.
[Statutory Authority: RCW 70.98.050 and 70.98.080. 09-06-003, § 246-240-113, filed 2/18/09, effective 3/21/09. Statutory Authority: RCW 70.98.050. 06-05-019, §
246-240-113, filed 2/6/06, effective 3/9/06.]