WAC 246-228-030
Facility requirements. (1) Radiation
levels. The local components of an analytical X-ray system
shall be located and arranged and shall include sufficient
shielding or access control such that no radiation levels
exist in any area surrounding the local component group which
could result in a dose to an individual present therein in
excess of the dose equivalent limits given in WAC 246-221-060
of these regulations. For systems utilizing X-ray tubes,
these levels shall be met at any specified tube rating.
(2) Surveys. Radiation surveys, as required by WAC 246-221-110 of all analytical X-ray systems, sufficient to
show compliance with WAC 246-228-030(1), shall be performed:
(a) Upon installation of the equipment, and at least once
every twelve months thereafter;
(b) Following any change in the initial arrangement,
number, or type of local components in the system;
(c) Following any maintenance requiring the disassembly
or removal of a local component in the system;
(d) During the performance of maintenance and alignment
procedures if the procedures require the presence of a primary
X-ray beam when any local component in the system is
disassembled or removed;
(e) Any time a visual inspection of the local components
in the system reveals an abnormal condition; and
(f) Whenever personnel monitoring devices required in WAC 246-228-050(2) show a significant increase over the previous
monitoring period or the readings are approaching 1/10 of the
hands and forearm limit specified in WAC 246-221-010.
(g) Radiation survey measurements shall not be required
if a registrant or licensee can demonstrate compliance to the
satisfaction of the department with WAC 246-228-030(1) in some
other manner.
(3) Posting. Each area or room containing analytical
X-ray equipment shall be conspicuously posted with a sign or
signs bearing the radiation symbol and the words "CAUTION - X-RAY
EQUIPMENT," or words having a similar intent.
(4) Documentation of instruction. Each facility shall
maintain written documentation showing that compliance with
WAC 246-228-050 has been met, and shall make such
documentation available to the department upon request.
[Statutory Authority: RCW 70.98.050 and 70.98.080. 91-15-083
(Order 183), § 246-228-030, filed 7/23/91, effective 8/23/91. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121),
recodified as § 246-228-030, filed 12/27/90, effective
1/31/91. Statutory Authority: RCW 70.98.050. 81-01-011
(Order 1570), § 402-40-040, filed 12/8/80; Order 1084, §
402-40-040, filed 1/14/76.]