WAC 246-225A-080
Special requirements for dental
intra-oral radiography. (1) Registrants using an X-ray system
designed for use with an intra-oral image receptor shall use
equipment that:
(a) Limits the SSD to not
less than 18 centimeters;
(b) Limits the X-ray beam so that the beam diameter at
the minimum SSD is no greater than 7 centimeters in diameter;
(c) Has an open-ended position-indicating device; and
(d) Has shielding included in the beam-limiting device or
position-indicating device equivalent to that required for the
diagnostic source assembly under WAC 246-225A-060(1).
(2) Registrants shall not use
diagnostic dental X-ray systems with a fixed, nominal kVp of less than 55.
[Statutory Authority: RCW 70.98.050 and 70.98.080. 11-19-013, § 246-225A-080, filed 9/7/11, effective 10/8/11;