WAC 246-225A-070
Special requirements for dental
extra-oral radiography. Registrants shall use X-ray systems
for extra-oral radiography that meet the following
requirements for:
(1) Beam limitation.
(a) X-ray equipment designed for only one image receptor
size at a fixed SID must be able to limit the size of the beam
at the plane of the image receptor to no larger than the image
receptor, and to align the center of the X-ray beam with the
center of the image receptor to within two percent of the SID.
In the case of extra-oral imaging systems where the image
receptor can be turned vertically or horizontally, the
beam-limiting device must also be able to be turned so that
the dimensions of the beam match the image receptor dimensions
at the image receptor plane.
(b) Intra-oral radiography systems used to perform
cephalometric projections, including trans-cranial exams, must
be equipped with a stable means to:
(i) Comply with the beam size dimensions in subsection
(1)(a) of this section; and
(ii) Center the beam to the image receptor as required in
subsection (1)(a) of this section.
(c) General purpose medical X-ray equipment used to
perform dental radiography must:
(i) Have stepless adjustment of the dimensions of the
X-ray beam so that the width and height of the X-ray beam are
independently adjustable. The minimum beam size at an SID of
100 centimeters must be equal to or less than 10 by 10
centimeters.
(ii) Have a means for operators to visually set the width
and height of the X-ray beam. The misalignment of the edges
of the visually set light field with the respective edges of
the X-ray beam along either the length or width of the
visually set light field must not be more than two percent of
the distance from the source to the center of the visually
defined light field when the surface upon which it appears is
perpendicular (at a 90 degree angle) to the central axis of
the X-ray beam.
(iii) Have a means to indicate on the X-ray equipment
when the axis of the X-ray beam is perpendicular to the plane
of the image receptor and to align the center of the X-ray
beam to the center of the image receptor to within two percent
of the SID (five percent for equipment manufactured before
August 1974). Dental lateral jaw examinations are excluded
from this requirement.
(iv) Have a beam-limiting device that indicates the X-ray
beam size in centimeters or inches at the plane of the image
receptor to which the beam-limiting device is adjusted.
(v) Have an actual beam size at the plane of the image
receptor that matches the indicated size to within two percent
of the SID.
(2) SSD and SID.
(a) Dental extra-oral radiography systems must have a
durable, securely fastened means to limit the SSD to not less
than 23 centimeters. The requirement may be met when the
beam-limiting device provides the required limits.
(b) Dental extra-oral radiography systems in which the
SID is not fixed (such as an intra-oral system used for
cephalometrics) must have a device or reference that will
indicate the actual SID to within two percent of the indicated
SID.
(3) Viewing device.
Dental extra-oral radiography installations must provide
a viewing device (mirror, video camera, or glass window
designed to reduce exposure) so that operators of the X-ray
equipment may observe the patient's head and neck area during
the exposure without being exposed to the primary beam or
stray radiation.
(4) Scattered radiation suppressing grids. When using
scattered radiation suppressing grids, the grids must be:
(a) Clearly labeled with the SID for which the grids are
designed to be used; and
(b) Used at the proper SID.
(5) X-ray film and screen requirements.
(a) X-ray film used for extra-oral imaging must be used
before the expiration date specified by the manufacturer.
(b) The spectral sensitivity of the X-ray film used must
be matched by the appropriate spectral output of the
intensifying screens used in the cassettes as recommended by
the film and screen manufacturers.
(c) Screens must be free of dirt, abrasions, and
discoloration that would cause artifacts on the image.
[Statutory Authority: RCW 70.98.050 and 70.98.080. 11-19-013, § 246-225A-070, filed 9/7/11, effective 10/8/11;
08-14-074, § 246-225A-070, filed 6/26/08, effective 7/27/08.]