| No grid |
= |
1.5 milliGray (100 millirads)/projection |
| Grid |
= |
2.5 milliGray (300 millirads)/projection |
| Xerox |
= |
4.0 milliGray (400 millirads)/projection |
(5) A quality control program shall be written and
implemented for all mammographic facilities. This shall
include (but shall not be limited to) tests performed, testing
frequency, testing protocol, control limits for each test,
corrective actions taken, and equipment maintenance/service. Program requirements include: