WAC 246-225-070
Radiographic systems other than
fluoroscopic, dental intraoral, or veterinary
systems -- Radiation exposure control devices. (1) Timers. Means shall be provided to terminate the exposure at a preset
time interval, preset product of current and time, a preset
number of pulses, or a preset radiation exposure to the image
receptor. In addition, it shall be impossible to make an
exposure when the timer is set to a zero or off position if
either position is provided.
(2) X-ray control (exposure switch):
(a) A control which shall be the equivalent of a dead-man
switch, shall be incorporated into each X-ray system such that
an exposure can be terminated at any time except for:
(i) Exposure of one-half second or less, or
(ii) During serial radiography when means shall be
provided to permit completion of any single exposure of the
series in process.
(b) Each X-ray control shall be located in such a way as
to meet the following requirements:
(i) Stationary X-ray systems shall be required to have
the X-ray exposure switch permanently mounted in a protected
area so that the operator has no choice but to remain in that
protected area during the entire exposure;
(ii) Mobile and portable X-ray systems shall have:
(A) An exposure cord which can extend for a minimum of 12
feet from the patient; or
(B) A protective barrier of 0.25 millimeter lead
equivalent between the patient and the operator.
(c) Each X-ray control shall provide visual evidence to
the operator that X rays are being produced and an audible
signal that the exposure has terminated.
(3) Automatic exposure controls (phototimers). When an
automatic exposure control is provided:
(a) Indication shall be made on the control panel when
this mode of operation is selected;
(b) When the X-ray tube potential is equal to or greater
than 50 kVp, the minimum exposure time for field emission
equipment rated for pulsed operation shall be equal to or less
than the interval equivalent to two pulses;
(c) The minimum exposure time for all equipment other
than that specified in WAC 246-225-070 (3)(b) shall be equal
to or less than 1/60 second or a time interval required to
deliver 5 mAs, whichever is greater.
(4) Timer reproducibility. With a timer setting of 0.5
seconds or less, the difference between the maximum exposure
time (Tmax) and the minimum exposure time (Tmin) shall be less
than or equal to 10% of the average exposure time (T), when
four timer tests are performed:
[Statutory Authority: RCW 70.98.050 and 70.98.080. 91-15-083
(Order 183), § 246-225-070, filed 7/23/91, effective 8/23/91. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121),
recodified as § 246-225-070, filed 12/27/90, effective
1/31/91. Statutory Authority: RCW 70.98.080. 83-19-050
(Order 2026), § 402-28-052, filed 9/16/83. Statutory
Authority: RCW 70.98.050. 81-01-011 (Order 1570), §
402-28-052, filed 12/8/80; Order 1084, § 402-28-052, filed
1/14/76. Formerly WAC 402-28-050 (part).]