WAC 246-225-060
Radiographic systems other than
fluoroscopic, dental intraoral, or veterinary systems -- Beam
limitation. The useful beam shall be limited to the area of
clinical interest and show evidence of collimation. This
shall be deemed to have been met if a positive beam limiting
device has been properly used or if evidence of collimation is
shown on at least three sides or three corners of the film,
(for example, projections from the shutters of the collimator,
cone cutting at the corners or a border at the film's edge).
(1) General purpose stationary and mobile X-ray systems.
(a) There shall be provided a means for stepless
adjustment of the size of the X-ray field such that at least
two dimensions of the X-ray field are independently variable. The minimum field size at a SID of 100 centimeters shall be
equal to or less than ten by ten centimeters.
(b) Adequate means shall be provided for visually
defining the perimeter of the X-ray field. The total
misalignment of the edges of the visually defined field with
the respective edges of the X-ray field along either the
length or width of the visually defined field shall not exceed
2 percent of the distance from the source to the center of the
visually defined field when the surface upon which it appears
is perpendicular to the central axis of the X-ray beam.
(2) In addition to the requirements of WAC 246-225-060(1)
above all stationary X-ray systems shall meet the following
requirements:
(a) Means shall be provided to indicate when the axis of
the X-ray beam is perpendicular to the plane of the image
receptor and to align the center of the X-ray field with
respect to the center of the image receptor to within 2
percent (5 percent for equipment manufactured prior to August
1974) of the SID. Dental lateral jaw examinations shall be
excluded from this requirement;
(b) The beam-limiting device shall numerically indicate
the field size in the plane of the image receptor to which it
is adjusted;
(c) Indication of field size dimensions and SID's shall
be specified in inches and/or centimeters;
(d) Indication of field size dimensions shall be such
that aperture adjustments result in X-ray field dimensions in
the plane of the image receptor to within 2 percent of the SID
when the beam axis is perpendicular to the plane of the image
receptor.
(3) Radiographic equipment designed for only one image
receptor size at a fixed SID shall be provided with means to
limit the field at the plane of the image receptor to
dimensions no greater than those of the image receptor, and to
align the center of the X-ray field with the center of the
image receptor to within 2 percent of the SID.
(4) Special purpose X-ray systems.
(a) These systems shall be provided with means to limit
the X-ray field in the plane of the image receptor so that
such field does not exceed each dimension of the image
receptor by more than 2 percent of the SID when the axis of
the X-ray beam is perpendicular to the plane of the image
receptor.
(b) These systems shall be provided with means to align
the center of the X-ray field with the center of the image
receptor to within 2 percent (5 percent for equipment
manufactured prior to August 1974) of the SID.
(c) The above WAC 246-225-060 (4)(a) and 246-225-060
(4)(b) may be met with a system that meets the requirements
for a general purpose X-ray system as specified in WAC 246-225-060(1) or, when alignment means are also provided, may
be met with either:
(i) An assortment of removable, fixed-aperture,
beam-limiting devices sufficient to meet the requirement for
each combination of image receptor size and SID for which the
unit is designed (each such device shall have clear and
permanent markings to indicate the image receptor size and SID
for which it is designed); or
(ii) A beam-limiting device having multiple fixed
apertures sufficient to meet the requirement for each
combination of image receptor size and SID for which the unit
is designed. Permanent, clearly legible markings shall
indicate the image receptor size and SID for which each
aperture is designed and shall indicate which aperture is in
position for use.
[Statutory Authority: RCW 70.98.050 and 70.98.080. 91-15-083
(Order 183), § 246-225-060, filed 7/23/91, effective 8/23/91. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121),
recodified as § 246-225-060, filed 12/27/90, effective
1/31/91. Statutory Authority: RCW 70.98.080. 83-19-050
(Order 2026), § 402-28-051, filed 9/16/83. Statutory
Authority: RCW 70.98.050. 81-01-011 (Order 1570), §
402-28-051, filed 12/8/80; Order 1084, § 402-28-051, filed
1/14/76. Formerly WAC 402-28-050 (part).]