WAC 246-243-120
Inspection and maintenance of radiographic
exposure devices, transport and storage containers, associated
equipment, source changers, and survey instruments. (1) The
licensee shall perform visual and operability checks on survey
meters, radiographic exposure devices, transport and storage
containers, associated equipment and source changers before use
on each day the equipment is to be used to ensure that the
equipment is in good working condition, that the sources are
adequately shielded, and that required labeling is present.
Survey instrument operability must be performed using check
sources or other appropriate means. If equipment problems are
found, the equipment must be removed from service until repaired.
(2) Each licensee shall have written procedures for:
(a) Inspection and routine maintenance of radiographic
exposure devices, source changers, associated equipment,
transport and storage containers, and survey instruments at
intervals not to exceed three months or before the first use
thereafter to ensure the proper functioning of components
important to safety. Replacement components shall meet design
specifications. If equipment problems are found, the equipment
must be removed from service until repaired.
(b) Inspection and maintenance necessary to maintain the
Type B packaging used to transport radioactive materials. The
inspection and maintenance program must include procedures to
assure that Type B packages are shipped and maintained in
accordance with the certificate of compliance or other approval.
(3) Any maintenance performed on radiographic exposure
devices and accessories shall be in accordance with the
manufacturer's specifications.
(4) Records of daily checks and quarterly inspections
including any equipment problems identified and of any
maintenance performed under subsections (1) and (2) of this
section shall be made and retained for three years. The record
shall include:
(a) The date of check or inspection;
(b) Name of inspector;
(c) Equipment involved;
(d) Any problems found; and
(e) What repair and/or maintenance, if any, was done.
[Statutory Authority: RCW 70.98.050. 00-08-013, § 246-243-120,
filed 3/24/00, effective 4/24/00; 94-01-073, § 246-243-120, filed
12/9/93, effective 1/9/94. Statutory Authority: RCW 70.98.050
and 70.98.080. 91-15-112 (Order 184), § 246-243-120, filed
7/24/91, effective 8/24/91. Statutory Authority: RCW 43.70.040.
91-02-049 (Order 121), recodified as § 246-243-120, filed
12/27/90, effective 1/31/91. Statutory Authority: RCW 70.98.080. 83-19-050 (Order 2026), § 402-36-095, filed 9/16/83. Statutory Authority: RCW 70.98.050. 81-01-011 (Order 1570), §
402-36-095, filed 12/8/80; Order 1084, § 402-36-095, filed
1/14/76.]